I’ve decided to split my exploration of dynamic arterial elastance into two articles. In this first article, we’re going to get a little geeky, exploring the physics and mathematics that underpin Eadyn. I’ve spent several days working with frontier AI models as mathematical research tools, constructing and interrogating progressively more sophisticated models of the circulation, isolating individual variables, testing hypotheses and challenging the physiological explanations in the published literature.
The models are necessarily simplified, but they allow us to test physiological explanations against the mathematics rather than relying on intuition alone. Although some of the concepts are quite complex, I’ve endeavoured to explain them as simply as possible, building the physiology step by step with worked numerical examples.
This is the sort of deep dive I want to offer paid subscribers to The Dependent Variable. The second article, which will be freely available, will draw on what we’ve established here to examine what Eadyn actually tells us and whether it’s clinically useful. You won’t need to have read this first article to understand the second, but for those who want to understand the physics behind the clinical conclusions, read on.
Stroke volume and pulse pressure are closely related. Each time the left ventricle ejects blood, it adds volume to the arterial circulation. The arteries distend, arterial pressure rises, and a pressure pulse travels through the arterial tree.
But the relationship between the volume ejected and the pressure pulse it produces is not constant.
A 10% increase in stroke volume might produce a 10% increase in pulse pressure. It might produce a 15% increase, or only 5%. The response depends on the properties of the arterial system, but also on the behaviour of ventricular ejection and what happened during preceding heartbeats.
During positive-pressure ventilation, the cyclic changes in ventricular loading cause stroke volume to vary. As we’ve discussed previously, this gives us stroke volume variation (SVV) and pulse pressure variation (PPV).
While these measurements are related, they aren’t necessarily equal.
Dynamic arterial elastance, or Eadyn, is simply their ratio:
Eadyn = PPV / SVV
If SVV is 10% and PPV is 15%, Eadyn is 1.5. Pulse pressure is varying proportionally more than stroke volume.
If SVV is 10% and PPV is 5%, Eadyn is 0.5. Pulse pressure is varying proportionally less than stroke volume.
If the two variations are equal, Eadyn is 1.
These are observations. We haven’t established anything about arterial stiffness, peripheral resistance, ventricular contractility or the response to treatment.
Yet Eadyn has been proposed as a measure of arterial loading, a marker of ventricular–arterial coupling, a predictor of the arterial pressure response to fluid administration, and a guide to withdrawing vasopressors.
We’ll examine those interpretations and their supporting evidence in the next article. First, we need to understand what determines the ratio itself. Why does a given variation in stroke volume sometimes produce a larger or smaller proportional variation in pulse pressure?
To answer that, we’ll build an arterial circulation from first principles, starting with a simple elastic chamber and progressively introducing the physical processes that influence the arterial pressure pulse.
The simplest possible artery
Imagine the entire arterial circulation as a single elastic chamber.
The left ventricle ejects blood into this chamber. As its volume increases, its walls stretch and the internal pressure rises.
For the moment, we’ll assume that no blood leaves during ejection and that the chamber has constant compliance.
Compliance describes how much additional volume the chamber accommodates for a given increase in pressure.
C = ΔV / ΔP
Compliance = change in volume / change in pressure
Rearranging:
ΔP = ΔV / C
Change in pressure = change in volume / compliance
If the entire stroke volume enters the chamber instantaneously, the pressure rise is:
PP = SV / C
Pulse pressure = stroke volume / compliance
Here PP represents the pressure jump produced by the stroke. In this idealised model, it is also the pulse pressure because pressure rises instantaneously and then decays between beats.
Let’s use actual numbers.
Suppose arterial compliance is 2 mL/mmHg. Every additional 2 mL accommodated by the arterial compartment increases its pressure by 1 mmHg.
The ventricle ejects 60 mL.
PP = 60 / 2 = 30 mmHg
Now imagine the next stroke volume is 66 mL.
PP = 66 / 2 = 33 mmHg
Stroke volume has increased by 10%, relative to the first stroke, and pulse pressure has also increased by 10%.
Because both have been divided by the same compliance, their proportional relationship is preserved.
Strictly, the conventional calculation of respiratory variation uses the midpoint of the maximum and minimum values, rather than the smaller value as its denominator.
For stroke volume:
SVV = (66 − 60) / 63 × 100 = 9.52%
For pulse pressure:
PPV = (33 − 30) / 31.5 × 100 = 9.52%
Both variations are equal, so Eadyn is 1.
Now suppose we repeat the experiment using an arterial chamber with half the compliance.
Compliance is now 1 mL/mmHg.
The 60 mL stroke produces a pulse pressure of 60 mmHg, while the 66 mL stroke produces a pulse pressure of 66 mmHg.
The chamber is twice as stiff. The absolute pulse pressures have doubled.
But PPV remains 9.52%, SVV remains 9.52%, and Eadyn remains 1.
The absolute compliance of the chamber has cancelled out of the calculation.
This is our first useful result. In this simplified model, arterial stiffness does not determine Eadyn. An extremely compliant arterial system and an extremely stiff arterial system can produce exactly the same ratio.
Of course, real arteries are not simple chambers with fixed compliance. As we progressively introduce the other physical properties of the circulation, some rather more surprising findings emerge. We’ll see how Eadyn can change substantially without any change in arterial compliance or resistance, and why the timing of ventricular ejection and even the behaviour of preceding heartbeats influence the result.